校正设备时,FID信号信号一条横直线/或者很乱

校正设备时,FID信号信号一条横直线/或者很乱

  ●首先确认射频单元是否打开,确保射频单元打开状态;

  ●查看探头内是否放置的是校准样品;

  ●查看主频SF及O1是否与出厂报告中或数据查询中之前数据正常时SF及O1是一致,不一致则修改为一致;

  ●观察(核磁共振)设备后面的接线是否接触不良,换上备用信号线。